The Pico Technology TA344 is a standard-grade manual SOLT calibration kit for accurate two-port vector network analyser calibration from 300 kHz to 8.5 GHz. It combines 50 Ω SMA male short, open, load and through standards in a four-piece, five-port assembly. Serial-number-specific traceable calibration data, tight-tolerance stainless steel interfaces and a protective carrying case support dependable VNA calibration for RF development, production testing and laboratory measurements..
The Standard 8.5 GHz SOLT Calibration Kit with SMA Male Connectors, order code TA344, is a manually operated mechanical calibration kit for PicoVNA and compatible 50 Ω vector network analysers.
The kit includes:
Pico supplies the standards as an assembled five-port SOLT unit. The components can be separated for individual use, connector inspection, cleaning or refurbishment if one standard becomes damaged.
SOLT calibration uses four known reference conditions:
During calibration, each mechanical standard is connected manually when requested by the VNA software.
The calibration allows the analyser to calculate corrections for systematic measurement errors such as:
The TA344 is not an electronic calibration module. It should not be described as USB-controlled, automatically switched or oven stabilised.
The TA344 provides SMA male interfaces across its short, open, load and through standards.
It is suitable for measurement setups terminating in:
The corresponding standard-grade female kit is the TA345.
The TA344 should not be described as having mixed male and female calibration ports. All five ports in the assembled SOLT kit are SMA male.
The TA344 is characterised from 300 kHz to 8.5 GHz.
This frequency range supports VNA calibration for measurements involving:
The full range should be shown as 300 kHz to 8.5 GHz, rather than only stating 8.5 GHz.
Pico describes the TA344 as a four-piece, five-port SOLT kit.
The assembly consists of:
Together, these provide five SMA male RF connection ports.
The modular format allows the standards to be:
This construction is practical for laboratories where the standards are used regularly.
Each TA344 kit is supplied with traceable calibration data linked to its individual serial number.
The data characterises the real frequency-dependent response of the:
This allows the VNA to compensate for non-ideal electrical behaviour, including:
The calibration data supplied for the individual TA344 should be used instead of generic or ideal standard definitions.
Pico offers standard-grade and premium-grade manual SOLT kits.
| Feature | TA344 Standard Kit | TA346 Premium Kit |
|---|---|---|
| Connector type | SMA male | PC3.5 male |
| Grade | Standard | Premium |
| Frequency range | 300 kHz to 8.5 GHz | 300 kHz to 8.5 GHz |
| Calibration type | Manual SOLT | Manual SOLT |
| Construction | Four-piece, five-port | Four-piece, five-port |
| Characterisation | Serial-number-specific | Serial-number-specific |
| Typical use | Working or transfer calibration | Precision or reference calibration |
Pico notes that calibration standards can be a major contributor to overall VNA measurement uncertainty. Some laboratories therefore use premium standards for reference work and standard-grade standards for regular calibration and transfer measurements.
The TA344 load standard provides the following published return-loss performance:
| Frequency Range | Uncorrected Return Loss | Corrected Return Loss |
|---|---|---|
| Up to 3 GHz | ≥28 dB | ≥40 dB |
| Above 3 GHz to 8.5 GHz | ≥26 dB | ≥37 dB |
Corrected return loss applies when the supplied serial-number-specific characterisation data is used during calibration.
The characterised load contributes to correction of:
The open standard has the following published return-loss limits:
| Frequency Range | Return Loss |
|---|---|
| Up to 3 GHz | ≤0.15 dB |
| Above 3 GHz to 8.5 GHz | ≤0.20 dB |
The open standard is not treated as an ideal infinite impedance. Its frequency-dependent capacitance, delay and connector response are represented in the supplied characterisation data.
The short standard provides:
| Frequency Range | Return Loss |
|---|---|
| Up to 3 GHz | ≤0.20 dB |
| Above 3 GHz to 8.5 GHz | ≤0.25 dB |
The supplied data accounts for the real delay, loss and inductive behaviour of the short standard.
The through adaptor provides:
It is used to calculate forward and reverse transmission corrections during two-port SOLT calibration.
A normal two-port calibration process includes:
The kit corrects systematic VNA errors, but final measurement quality also depends on test cables, connector condition, torque, cable movement and environmental stability.
SMA calibration standards require careful handling.
Recommended practices include:
Connector wear, contamination or incorrect torque can increase measurement uncertainty even when the correct calibration data is applied.
A characterised check standard can be measured after calibration to verify the complete measurement setup.
Compatible Pico check standards include:
Verification can help identify:
The TA344 establishes the VNA error correction, while a check standard independently evaluates the corrected measurement setup.
The TA344 can be used with:
The selected sweep range must remain within the kit’s 300 kHz to 8.5 GHz characterisation range.
The Pico TA344 is suitable for:
Includes mechanical:
Supports calibration across low-frequency RF, wireless and microwave applications.
Corrected return loss:
Supports safe storage and transport of the calibration standards.
| Specification | Pico TA344 |
|---|---|
| Product name | Standard 8.5 GHz SOLT Calibration Kit, SMA Male |
| Order code | TA344 |
| Product type | Standard-grade manual mechanical VNA calibration kit |
| Calibration method | Two-port SOLT |
| Standards included | Short, open, load and through |
| Construction | Four-piece, five-port assembly |
| Connector type | SMA male |
| Nominal impedance | 50 Ω |
| Characterised frequency range | 300 kHz to 8.5 GHz |
| Connector construction | Tight-tolerance stainless steel |
| Characterisation | Serial-number-specific traceable calibration data |
| Electronic switching | No |
| USB interface | No |
| USB power | No |
| Internal temperature oven | No |
| Protective carrying case | Included |
| Parameter | Up to 3 GHz | Above 3 GHz to 8.5 GHz |
|---|---|---|
| Uncorrected return loss | ≥28 dB | ≥26 dB |
| Corrected return loss | ≥40 dB | ≥37 dB |
| Parameter | Up to 3 GHz | Above 3 GHz to 8.5 GHz |
|---|---|---|
| Return loss | ≤0.15 dB | ≤0.20 dB |
| Parameter | Up to 3 GHz | Above 3 GHz to 8.5 GHz |
|---|---|---|
| Return loss | ≤0.20 dB | ≤0.25 dB |
| Specification | Value |
|---|---|
| Maximum insertion loss | 0.25 dB |
| Maximum characterised frequency | 8.5 GHz |
| Connector configuration | SMA male to SMA male |
| Nominal impedance | 50 Ω |
| Characterisation data | Included |
| Specification | Details |
|---|---|
| Data type | Serial-number-specific calibration characterisation |
| Frequency range | 300 kHz to 8.5 GHz |
| Standards characterised | Short, open, load and through |
| Traceability | Traceable calibration data supplied |
| Typical use | PicoVNA and compatible VNA calibration |
| Electronic switching | Not applicable |